Bacharach ECA 450 NOx Kit with NO / NO2 sensors and compact sample conditioner

Bacharach ECA 450 NOx Kit with NO / NO2 sensors and compact sample conditioner - Bacharach ECA 450

Bacharach ECA 450 NOx Kit with NO / NO2 sensors and compact sample conditioner

Item Id: 0024-8400
Your Price:
$9,015.00
Availability: 3-5 days delivery Call for Quote.

Quantity

Description

The ECA 450 is ideal for professionals concerned about combustion efficiency, environmental compliance, or both. It enables plant maintenance engineers and managers, industrial boiler/furnace service technicians, energy coordinators, compliance officers, environmental auditors and safety managers ensure that industrial equipment is burning efficiently while environmental regulations are being met.

Unlike any other industrial grade combustion efficiency and environmental analyzer on the market today, the new ECA 450 from Bacharach conducts accurate combustion and emissions tests that will help you meet environmental regulations and ensure your equipment is burning efficiently. The ECA 450 measures Oxygen, Carbon Monoxide, Nitric Oxide, Nitrogen Dioxide and Sulfur Dioxide, as well as combustibles, temperature and draft. It also calculates combustion efficiency, excess air, carbon dioxide, NOx (Nitrogen Oxides) and pollution units.

The ECA takes continuous samplings of furnaces, boilers and other industrial combustion equipment for up to eight hours and generates precise, reliable combustion and environmental measurements and calculations. And it generates this complex data easily even at selected intervals enabling plants to make more accurate equipment adjustments and better assess pollution levels. The ECA 450 helps improve overall combustion efficiency, reduce fuel costs, and determine emissions compliance.


  • Large, easy-to-read fluorescent display that is visible from wide angles.
  • Battery life of over eight hours.
  • Modular design lets you add sensors as your needs change.
  • Built-in printer and complete data downloading to a PC capability.
  • EPA-ETV test verified performance.
  • Measures and displays O2 and CO in the flue gas, pressure (draft), primary air temperature, and stack temperature.
  • Calculates and displays combustion efficiency, excess air, and CO2.
  • Optionally measures and displays NO, NO2, SO2, HC (methane equivalent combustibles), and CO in the high range of 4,000 to 80,000 ppm).
  • Optionally calculates NOx (the combination of NO and NO2), and calculates CO, SO2, and NOx individually referenced to a user defined Oxygen level of between 0 and 15%.
  • Optionally displays pollution conversions for CO, NO, NO2, and SO2. Pollution conversions include parts per million, pounds of pollutant per million BTU, milligrams of pollutant per cubic meter of gas, and grams of pollutant per gigajoule.
  • Stores over 1000 individual combustion and/or pressure test records, which can later be recalled for viewing or printing.
  • Stored test records can be downloaded to a personal computer and stored as a text file, which can then be opened in a spreadsheet program for analysis.
  • Performs data logging.


Specifications:


Measured Data
Oxygen0.1 to 20.9%
Carbon Monoxide (hydrogen compensated)0 to 4,000 ppm
Carbon Monoxide High4,001 to 80,000 ppm
Nitric Oxide*0 to 3,500 ppm
Nitrogen Dioxide*0 to 500 ppm
Sulfur Dioxide*0 to 4,000 ppm
Combustibles0 to 5% (application dependent)
Stack Temp.-4 to 2400F (-20 to 1315C)
Primary/Ambient Temp.-4 to 999F (-20 to 999C)
Pressure/Draft-27.7 to 27.7 inches of Water

Calculated Data
Combustion Efficiency0.1 to 100.0%
Excess Air1.0 to 250%
Carbon Dioxide(dry basis)0 to fuel dependent maximum
NOx0 to 4,000 ppm
NOx (ref. to %O2)0 to 17,000 ppm
CO (ref. to %O2)0 to 99,9999 ppm
NO (ref. to %O2)0 to 14,900 ppm
NO2 (ref. to %O2)0 to 2100 ppm
SO2 (ref. to %O2)0 to 17,000 ppm

Miscellaneous
PowerUniversal AC adapter and an internal battery pack. Adapter will accept input voltages from 100 to 240V. Fully charged battery pack provides a minimum of 8 hours of operation

Pumps & Probe

Two pumps are included. The first pump supplies gas sample to the sensors. The second pump supplies fresh air to purge the low-range CO sensor when CO levels exceed 400 ppm. Probe includes a standard probe and hose assembly equipped with a water trap, particulate filter, probe stop, 15 feet of hose, and 12-inch probe tube

Size

13.5"H x 18.5"W x 9"D (342.9 x 467.0 x 228.6 mm)

Weight

25 lbs (11.34 kg)

Warranty

1 year, extended warranty available