Bacharach 6600-8218 Propane (0 to 100% LEL) Infrared Sensor for MGS-550
Measuring Range: 0 - 100%
Replacing the Sensor Module
- Power down the instrument.
- Loosen the set screw (see photos to right) in the sensing head cap (item 8) using a 1.5 mm hex wrench.
- Unscrew the sensing head cap (item 8) and remove.
- Pull out sensor module (item 5) from sensor control board (item 4).
- With flat edges of PCBs aligned, plug new sensor module into sensor control board.
- Screw on the sensing head cap to the sensing head base (hand tight).
- Tighten the set screw in the sensing head cap.
- Power-up the instrument.
- Verify calibration.